HydroRad Spectrometer Options
blackrule
 
Standard Spectrometer
  • Calibrated Wavelength Range: 350 nm to 850 nm
  • Pixel Resolution: 0.30 to 0.35nm
  • Entrance aperture: 50 �m wide slit
  • Order sorting filters: Yes
  • Stray light: <0.05% at 600 nm; <0.10% at 435 nm
  • Fixed integration time dynamic range: 200:1
  • Variable integration time dynamic range: 5 x 105:1
  • Integration time: 21 milliseconds to 60 seconds
  • Irradiance sensitivity at 1 second integration: 10-5 W/m2/nm at 600 nm; 4x10-5 W/m2/nm at 435 nm
  • Radiance sensitivity at 1 second integration: 10-6 W/m2/nm/sr at 600 nm; 1.5x10-6 W/m2/nm/sr at 435 nm

Reflection Efficiency of Standard Spectrometer Grating

 

                                                                                                                      

 

Standard Near-Infrared Spectrometer
  • Calibrated Wavelength Range: 400 nm to 1000 nm
  • Pixel Resolution: 0.30 to 0.35nm
  • Entrance aperture: 50 �m wide slit
  • Order sorting filters: Optional
  • Stray light: <0.05% at 600 nm; <0.10% at 435 nm
  • Fixed integration time dynamic range: 200:1
  • Variable integration time dynamic range: 5 x 105:1
  • Integration time: 21 milliseconds to 60 seconds
  • Irradiance sensitivity at 1 second integration: 8x10-6 W/m2/nm at 600 nm; 4.5x10-5 W/m2/nm at 435 nm
  • Radiance sensitivity at 1 second integration: 8x10-7 W/m2/nm/sr at 600 nm; 1.7x10-6 W/m2/nm/sr at 435 nm 

Reflection Efficiency of Near-IR Spectrometer Grating

                                                                                                                      

 

High-resolution composite grating spectrometer
  • Calibrated Wavelength Range: 300 nm to 1100 nm
  • Pixel Resolution: 0.40 to 0.45nm
  • Entrance aperture: selectable 5 �m or 50 �m wide slit
  • Order sorting filters: Yes
  • Stray light: <0.05% at 600 nm; <0.10% at 435 nm
  • Fixed integration time dynamic range: 200:1
  • Variable integration time dynamic range: 4 x 106:1
  • Integration time: 3 milliseconds to 60 seconds
  • Irradiance sensitivity at 1 second integration: 1.4x10-6 W/m2/nm at 600 nm; 5x10-6 W/m2/nm at 435 nm
  • Radiance sensitivity at 1 second integration: 1.4x10-7 W/m2/nm/sr at 600 nm; 2x10-7 W/m2/nm/sr at 435 nm 

Reflection Efficiency of Composite Spectrometer Grating

                          

                                                                                                                         

Copyright © 1998-2002 by Hydro-Optics, Biology, & Instrumentation Laboratories
All rights reserved.
All specifications subject to change without notice.

販売 ワンピース dvd box 全話 golf 販売 golf 通販 Taylormade Callaway Ping 犬夜叉 DVD Windows OS 販売 Windows 7 Ultimate 天国への階段 トライガン 花より男子 のだめカンタービレ アタックNO.1 wholesale 60 Day Workout 60 Day Workout dvd 60 Day Workout Prison Break dvd The OFFICE dvd BUFFY THE VAMPIRE SLAYER dvd NCIS dvd Family Guy DVD The War DVD Girlfriends DVD NARUTO DVD American Pie 1-7